> CVE-2026-7251 [CRITICAL]

« BACK_TO_HOME
PUBLISHED
2026-05-26 18:16
MODIFIED
2026-05-26 18:16
CVSS_SCORE V3
9.3
EPSS (Prob. Exploit)
0.50%
Top 57.3%
DATA SOURCES:
NVD 🔗 2026-05-26 CWE MITRE 🔗 2026-11-01 FIRST EPSS 🔗 2026-11-01

> DESCRIPTION

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

> AFFECTED_PRODUCTS

ℹ️ Aucun produit affecté n'a encore été publié par NVD pour cette CVE.

> CVSS V3 METRICS

CVSS_VECTOR
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

> MÉTRIQUES CVSS DÉTAILLÉES

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Métrique Valeur Interprétation
AV (Vecteur d'Attaque)
Attack Vector
Network (N)
Exploit possible à distance via réseau (HTTP, API, etc.)
AC (Complexité d'Attaque)
Attack Complexity
Low (L)
Peu de conditions requises pour exploiter
PR (Privilèges Requis)
Privileges Required
None (N)
Aucune authentification requise
UI (Interaction Utilisateur)
User Interaction
None (N)
Aucune interaction utilisateur nécessaire
S (Périmètre)
Scope
Unchanged (U)
Impact reste dans le périmètre du composant compromis
C (Impact Confidentialité)
Confidentiality Impact
High (H)
Divulgation totale d'informations
I (Impact Intégrité)
Integrity Impact
High (H)
Modification totale des données possible
A (Impact Disponibilité)
Availability Impact
High (H)
Interruption totale du service possible

> RÉSUMÉ DES IMPACTS

Confidentialité
High
Divulgation totale d'informations
Intégrité
High
Modification totale des données possible
Disponibilité
High
Interruption totale du service possible

> WEAKNESSES

CWE-259

> CWE DETAILED DESCRIPTIONS

CWE-259: Use of Hard-coded Password - The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound authentication to external components.

> HISTORIQUE DES MODIFICATIONS

Première apparition
26/05/2026 18:16
Source: NVD
📊 ENRICHISSEMENT 01/11/2026 06:59
CWE
Champ modifié: Description CWE
Ancienne valeur:
CWE-259: Use of Hard-coded Password - The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.
Nouvelle valeur:
CWE-259: Use of Hard-coded Password - The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound authentication to external components.
📊 ENRICHISSEMENT 01/06/2026 06:01
CWE
Champ modifié: Description CWE
Ancienne valeur:
N/A
Nouvelle valeur:
CWE-259: Use of Hard-coded Password - The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.
Total des modifications: 2 Dernière mise à jour: 6 days ago

> REFERENCES

https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-1...
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01
https://www.eppendorf.com/software-downloads
« BACK_TO_HOME