> CVE-2026-66875 [HIGH]

« BACK_TO_HOME
PUBLISHED
2026-08-11 22:18
MODIFIED
2026-08-11 22:18
CVSS_SCORE V3
8.7
EPSS (Prob. Exploit)
0.40%
Top 65.2%
DATA SOURCES:
NVD đź”— 2026-08-11 CWE MITRE đź”— 2026-11-01 FIRST EPSS đź”— 2026-11-01

> DESCRIPTION

In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.

> AFFECTED_PRODUCTS

ℹ️ Aucun produit affecté n'a encore été publié par NVD pour cette CVE.

> CVSS V3 METRICS

CVSS_VECTOR
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

> MÉTRIQUES CVSS DÉTAILLÉES

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Métrique Valeur Interprétation
AV (Vecteur d'Attaque)
Attack Vector
Adjacent (A)
Exploit nécessite un accès au segment réseau local
AC (Complexité d'Attaque)
Attack Complexity
Low (L)
Peu de conditions requises pour exploiter
PR (Privilèges Requis)
Privileges Required
None (N)
Aucune authentification requise
UI (Interaction Utilisateur)
User Interaction
None (N)
Aucune interaction utilisateur nécessaire
S (Périmètre)
Scope
Unchanged (U)
Impact reste dans le périmètre du composant compromis
C (Impact Confidentialité)
Confidentiality Impact
High (H)
Divulgation totale d'informations
I (Impact Intégrité)
Integrity Impact
High (H)
Modification totale des données possible
A (Impact Disponibilité)
Availability Impact
High (H)
Interruption totale du service possible

> RÉSUMÉ DES IMPACTS

Confidentialité
High
Divulgation totale d'informations
Intégrité
High
Modification totale des données possible
Disponibilité
High
Interruption totale du service possible

> WEAKNESSES

CWE-306

> CWE DETAILED DESCRIPTIONS

CWE-306: Missing Authentication for Critical Functionality - The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

> HISTORIQUE DES MODIFICATIONS

Première apparition
11/08/2026 22:18
Source: NVD
📊 ENRICHISSEMENT 01/11/2026 06:59
CWE
Champ modifié: Description CWE
Ancienne valeur:
CWE-306: Missing Authentication for Critical Function - The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of res...
Nouvelle valeur:
CWE-306: Missing Authentication for Critical Functionality - The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount o...
📊 ENRICHISSEMENT 01/09/2026 06:02
CWE
Champ modifié: Description CWE
Ancienne valeur:
N/A
Nouvelle valeur:
CWE-306: Missing Authentication for Critical Function - The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of res...
Total des modifications: 2 Dernière mise à jour: 10 hours ago

> REFERENCES

https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-2...
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01
« BACK_TO_HOME