> CVE-2019-25720 [HIGH]

« BACK_TO_HOME
PUBLISHED
2026-06-03 18:16
MODIFIED
2026-06-03 18:16
CVSS_SCORE V3
7.1
EPSS (Prob. Exploit)
0.20%
Top 90.7%
DATA SOURCES:
NVD đź”— 2026-06-03 CWE MITRE đź”— 2026-11-01 FIRST EPSS đź”— 2026-10-25

> DESCRIPTION

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

> AFFECTED_PRODUCTS

ℹ️ Aucun produit affecté n'a encore été publié par NVD pour cette CVE.

> CVSS V3 METRICS

CVSS_VECTOR
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

> MÉTRIQUES CVSS DÉTAILLÉES

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Métrique Valeur Interprétation
AV (Vecteur d'Attaque)
Attack Vector
Adjacent (A)
Exploit nécessite un accès au segment réseau local
AC (Complexité d'Attaque)
Attack Complexity
Low (L)
Peu de conditions requises pour exploiter
PR (Privilèges Requis)
Privileges Required
None (N)
Aucune authentification requise
UI (Interaction Utilisateur)
User Interaction
None (N)
Aucune interaction utilisateur nécessaire
S (Périmètre)
Scope
Unchanged (U)
Impact reste dans le périmètre du composant compromis
C (Impact Confidentialité)
Confidentiality Impact
None (N)
Aucune divulgation d'informations
I (Impact Intégrité)
Integrity Impact
None (N)
Aucune modification de données possible
A (Impact Disponibilité)
Availability Impact
High (H)
Interruption totale du service possible

> RÉSUMÉ DES IMPACTS

Confidentialité
None
Aucune divulgation d'informations
Intégrité
None
Aucune modification de données possible
Disponibilité
High
Interruption totale du service possible

> WEAKNESSES

CWE-1286

> CWE DETAILED DESCRIPTIONS

CWE-1286: Improper Validation of Syntactic Correctness of Input - The product receives input that is expected to be well-formed - i.e., to comply with a certain syntax - but it does not validate or incorrectly validates that the input complies with the syntax.

> HISTORIQUE DES MODIFICATIONS

Première apparition
03/06/2026 18:16
Source: NVD
📊 ENRICHISSEMENT 01/07/2026 06:01
CWE
Champ modifié: Description CWE
Ancienne valeur:
N/A
Nouvelle valeur:
CWE-1286: Improper Validation of Syntactic Correctness of Input - The product receives input that is expected to be well-formed - i.e., to comply with a certain syntax - but it does not validate or in...
Total des modifications: 1 Dernière mise à jour: 4 months ago

> REFERENCES

https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-se...
https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malfor...
« BACK_TO_HOME