> CVE-2019-25719 [HIGH]

« BACK_TO_HOME
PUBLISHED
2026-06-02 14:16
MODIFIED
2026-06-02 14:40
CVSS_SCORE V3
8.8
EPSS (Prob. Exploit)
0.13%
Top 97.5%
DATA SOURCES:
NVD 🔗 2026-06-02 CWE MITRE 🔗 2026-11-01 FIRST EPSS 🔗 2026-10-25

> DESCRIPTION

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

> AFFECTED_PRODUCTS

ℹ️ Aucun produit affecté n'a encore été publié par NVD pour cette CVE.

> CVSS V3 METRICS

CVSS_VECTOR
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

> MÉTRIQUES CVSS DÉTAILLÉES

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Métrique Valeur Interprétation
AV (Vecteur d'Attaque)
Attack Vector
Network (N)
Exploit possible à distance via réseau (HTTP, API, etc.)
AC (Complexité d'Attaque)
Attack Complexity
Low (L)
Peu de conditions requises pour exploiter
PR (Privilèges Requis)
Privileges Required
None (N)
Aucune authentification requise
UI (Interaction Utilisateur)
User Interaction
None (N)
Aucune interaction utilisateur nécessaire
S (Périmètre)
Scope
Unchanged (U)
Impact reste dans le périmètre du composant compromis
C (Impact Confidentialité)
Confidentiality Impact
Low (L)
Divulgation limitée d'informations
I (Impact Intégrité)
Integrity Impact
Low (L)
Modification limitée des données possible
A (Impact Disponibilité)
Availability Impact
High (H)
Interruption totale du service possible

> RÉSUMÉ DES IMPACTS

Confidentialité
Low
Divulgation limitée d'informations
Intégrité
Low
Modification limitée des données possible
Disponibilité
High
Interruption totale du service possible

> WEAKNESSES

CWE-924

> CWE DETAILED DESCRIPTIONS

CWE-924: Improper Enforcement of Message Integrity During Transmission in a Communication Channel - The product establishes a communication channel with an endpoint and receives a message from that endpoint, but it does not sufficiently ensure that the message was not modified during transmission.

> HISTORIQUE DES MODIFICATIONS

Première apparition
02/06/2026 14:16
Source: NVD
📊 ENRICHISSEMENT 01/07/2026 06:01
CWE
Champ modifié: Description CWE
Ancienne valeur:
N/A
Nouvelle valeur:
CWE-924: Improper Enforcement of Message Integrity During Transmission in a Communication Channel - The product establishes a communication channel with an endpoint and receives a message from that en...
Total des modifications: 1 Dernière mise à jour: 4 months ago

> REFERENCES

https://static.draeger.com/security
https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-v...
« BACK_TO_HOME